IE4: Vital Signs Monitor

产品规格

product details

TECHNICAL SPECIFICATIONS

Quality Standards and Classification

CE, ISO13485

SFDA: Class Ⅱb

Anti-electroshock degree:

ClassⅠequipment

(internal power supply)

ECG/Resp: CF

Application Range

Adult/Pediatric/Neonatal/Medicine/Surgery/

Operating Room/ICU/CCU

Display

4” real color TFT screen

Resolution: 800*600

One alarm indicator(yellow/red)

One working indicator(green)

One battery charge state indicator(green)

Three modes in accordance with the alarm state

Environment

Operating environment :

Temperature: 0 ~ 40 ℃

Humidity: ≤85%

Altitude: -500 ~ 4600m

Power Requirements

AC: 100 ~ 240V, 50Hz/60Hz

DC: Built-in rechargeable battery

Battery: 4000mA 11.1V lithium battery

2h operating after full charge(one piece)

5min operating after low battery alarm

Date Storage

Trend diagram/table: 7x24h

NIBP review: 400 events

Wave review: 100min

Alarm review: 100 alarm events

Support drug concentration titration analysis

ST Segment

ST Segment Range: -2.0mV ~ +2.0mV

Accuracy: 0.02mV

ECG

5 Leads: RA, LA, LL, RL, V

Lead mode: Ⅰ,Ⅱ, Ⅲ, aVR, aVL, aVF, V

Increase: ×250,×500,×1000,×2000

Sensitivity: > 200 uV (Peak-to-peak value)

Input impedance: > 5 ( megohm )

Bandwidth: Surgery 1 ~ 20 Hz

Monitor 0.5 ~ 40 Hz

Diagnostic 0.05 ~ 130 Hz

CMRR: ≥ 100dB

Polarization Voltage: ± 300mV

Baseline Recovering Time: After defibrillation< 3 seconds

Signal Range: 8 mV (Peak-to-peak value)

Calibrating Signal: 1mV, precision ±5%

RESP

Method: RA-LL impedance

Resp Impedance range: 0.3 ~ 3 Ω

Base Impedance range: 200 Ω ~ 4000 Ω

Bandwidth: 0.1 ~ 2.5Hz

Resp rate: Adult 0 ~ 120BrPM

Neonatal/Pediatric 0 ~ 150BrPM

Resolution: 1BrPM

Precision: ±2 BrPM

Asphyxia Alarm: 10~40 seconds

NIBP

Method: Pulse wave oscillometry

Work mode: Manual/ Auto/ STAT

Measure interval of auto mode:

1,2,3,4,5,10,15,30,60,90,120,180,240,480 minute(s)

Measuring Time of STAT Mode: 5 minutes

PR range: 40 ~ 240bpm

Measure & alarm range: Adult

SYS 40 ~ 270mmHg

DIA 10 ~ 215mmHg

MEAN 20 ~ 235mmHg

Pediatric

SYS 40 ~ 200mmHg

DIA 10 ~ 150mmHg

MEAN 20 ~ 165mmHg

Neonatal

SYS 40 ~ 135mmHg

DIA 10 ~ 100mmHg

MEAN 20 ~ 110mmHg

Static pressure range: 0 ~ 300mmHg

Precision: ± 3mmHg

Pressure precision:

Max. average error: ±5mmHg

Max. standard deviation: ±8mmHg

Overvoltage protection:

Adult 300mmHg

Pediatric 240mmHg

Neonatal 150mmHg

SpO2

Range: 0 ~ 100%

Resolution: 1%

Precision: 70% ~ 100%: ±2 DIGIT

0% ~ 69%: ±no definition given

Pulse Rate

Range: 20 ~ 300bpm

Resolution: 1bpm

Precision: ±3bpm

TEMP

Channel 2

Measure & alarm range: 0 ~ 50 ℃

Resolution: 0.1 ℃

Precision (no sensor): ±0.1 ℃

Standard accessories

NIBP cuff & tube

ECG cable & electrodes

SpO2 sensor

TEMP probe

Lithium-ion battery

Power cable

Operator’s manual

Optional accessories

CO2 module

IBP module

Touch screen

Trolley bracket

Hanging bracket

Monitor recorder

IBP


Measurement range:

0-150mmHg, 0-19.7%,0-20kPa (at 760mmHg),

atmospheric pressure provided by the host.

Resolution 0.1mmHg : 0-69mmHg

0.25mmHg: 70-150mmHg

Precision 0-40mmHg: ±2mmHg

41-70mmHg: ±5%(reading)

71-100mmHg: ±8%(reading)

101-150mmHg:±10%(reading)

Respiratory rate range 0- 150 BPM

Respiratory rate accuracy: ±1 BPM