IE4: Vital Signs Monitor
product details
TECHNICAL SPECIFICATIONS
Quality Standards and Classification
CE, ISO13485
SFDA: Class Ⅱb
Anti-electroshock degree:
ClassⅠequipment
(internal power supply)
ECG/Resp: CF
Application Range
Adult/Pediatric/Neonatal/Medicine/Surgery/
Operating Room/ICU/CCU
Display
4” real color TFT screen
Resolution: 800*600
One alarm indicator(yellow/red)
One working indicator(green)
One battery charge state indicator(green)
Three modes in accordance with the alarm state
Environment
Operating environment :
Temperature: 0 ~ 40 ℃
Humidity: ≤85%
Altitude: -500 ~ 4600m
Power Requirements
AC: 100 ~ 240V, 50Hz/60Hz
DC: Built-in rechargeable battery
Battery: 4000mA 11.1V lithium battery
2h operating after full charge(one piece)
5min operating after low battery alarm
Date Storage
Trend diagram/table: 7x24h
NIBP review: 400 events
Wave review: 100min
Alarm review: 100 alarm events
Support drug concentration titration analysis
ST Segment
ST Segment Range: -2.0mV ~ +2.0mV
Accuracy: 0.02mV
ECG
5 Leads: RA, LA, LL, RL, V
Lead mode: Ⅰ,Ⅱ, Ⅲ, aVR, aVL, aVF, V
Increase: ×250,×500,×1000,×2000
Sensitivity: > 200 uV (Peak-to-peak value)
Input impedance: > 5 ( megohm )
Bandwidth: Surgery 1 ~ 20 Hz
Monitor 0.5 ~ 40 Hz
Diagnostic 0.05 ~ 130 Hz
CMRR: ≥ 100dB
Polarization Voltage: ± 300mV
Baseline Recovering Time: After defibrillation< 3 seconds
Signal Range: 8 mV (Peak-to-peak value)
Calibrating Signal: 1mV, precision ±5%
RESP
Method: RA-LL impedance
Resp Impedance range: 0.3 ~ 3 Ω
Base Impedance range: 200 Ω ~ 4000 Ω
Bandwidth: 0.1 ~ 2.5Hz
Resp rate: Adult 0 ~ 120BrPM
Neonatal/Pediatric 0 ~ 150BrPM
Resolution: 1BrPM
Precision: ±2 BrPM
Asphyxia Alarm: 10~40 seconds
NIBP
Method: Pulse wave oscillometry
Work mode: Manual/ Auto/ STAT
Measure interval of auto mode:
1,2,3,4,5,10,15,30,60,90,120,180,240,480 minute(s)
Measuring Time of STAT Mode: 5 minutes
PR range: 40 ~ 240bpm
Measure & alarm range: Adult
SYS 40 ~ 270mmHg
DIA 10 ~ 215mmHg
MEAN 20 ~ 235mmHg
Pediatric
SYS 40 ~ 200mmHg
DIA 10 ~ 150mmHg
MEAN 20 ~ 165mmHg
Neonatal
SYS 40 ~ 135mmHg
DIA 10 ~ 100mmHg
MEAN 20 ~ 110mmHg
Static pressure range: 0 ~ 300mmHg
Precision: ± 3mmHg
Pressure precision:
Max. average error: ±5mmHg
Max. standard deviation: ±8mmHg
Overvoltage protection:
Adult 300mmHg
Pediatric 240mmHg
Neonatal 150mmHg
SpO2
Range: 0 ~ 100%
Resolution: 1%
Precision: 70% ~ 100%: ±2 DIGIT
0% ~ 69%: ±no definition given
Pulse Rate
Range: 20 ~ 300bpm
Resolution: 1bpm
Precision: ±3bpm
TEMP
Channel 2
Measure & alarm range: 0 ~ 50 ℃
Resolution: 0.1 ℃
Precision (no sensor): ±0.1 ℃
Standard accessories
NIBP cuff & tube
ECG cable & electrodes
SpO2 sensor
TEMP probe
Lithium-ion battery
Power cable
Operator’s manual
Optional accessories
CO2 module
IBP module
Touch screen
Trolley bracket
Hanging bracket
Monitor recorder
IBP
Measurement range:
0-150mmHg, 0-19.7%,0-20kPa (at 760mmHg),
atmospheric pressure provided by the host.
Resolution 0.1mmHg : 0-69mmHg
0.25mmHg: 70-150mmHg
Precision 0-40mmHg: ±2mmHg
41-70mmHg: ±5%(reading)
71-100mmHg: ±8%(reading)
101-150mmHg:±10%(reading)
Respiratory rate range 0- 150 BPM
Respiratory rate accuracy: ±1 BPM
________________________________________________________________________________________________________________________________________________________________________________________________________